SUMASTIK 9050
Test speed: |
Thoughput 400test/hour,800 test with ISE |
Measuring wavelength: |
340 to 800mm, 12 wavelengths invariable |
Assay tape: |
End point, Kinetics, fixed time, immunoturbidimetry Monochromatic, |
Bi-chromatic&Multi-standard Analysis Linearity&Nonlinearity |
|
Calibration Single&Dual reagent can be used. |
|
Calibration Mode: |
Linear(one-point,Two-point and Multi-point),Logit-log4p, Logit-log5Spine,Exponential, Polynomial, Parabola. |
Light source: |
Halogen lamp 6V/10W, Life span≥3000hours |
Cuvette: |
Light pass length:6mm |
Reaction disk: |
Turntable |
Teaction temperature: |
37℃±0.1℃ |
Sample Volume: |
2 to 35μL(in steps of 0.1μL) |
Power supply: |
220V AC, 50/60HZ |
Copyright 長春賽諾邁德有限公司 版權(quán)所有.